Calibration process for KH keeper's pH probe

Calibration process for KH keeper's pH probe

Leave pH probe to soak in saltwater in your glass beaker for at least 30 minutes. Then please do as follows:

  1. Heat the pH 4.01 and pH 7.01 solution to 25.0 Celsius.

  2. Place the probe into the pH 4.01 solution. Leave it there for at least 5 minutes.

  3. Start calibration procedure for pH 4 solution.

  4. After 5 minutes, rinse in RO water, pat dry.

  5. Place into pH 7.01 solution. Leave it there for at least 5 minutes.

  6. Start calibration procedure for pH 7 solution.

  7. Complete the calibration.

After calibration please test accuracy.

Leave pH probe for 5 minutes in pH 4.01 solution and test accuracy. The difference shouldn’t be higher than 0.05.

Then leave the pH probe in pH 7.01 solution for 5 minutes to check the pH probe is accurate to the measurement solution. The difference shouldn’t be higher than 0.05.

Here is the video about the calibration:

Regards

Reef Factory Team

 