Leave pH probe to soak in salt water in your glass beaker for at least 30 minutes. Then please do as follows:

  1. Heat the pH 4.01 and pH 7.01 solution to 25.0 Celsius.

  2. Place the probe into the pH 4.01 solution. Leave it there for at least 5 minutes.

  3. Start calibration procedure for pH 4 solution.

  4. After 5 minutes, rinse in RO water, pat dry.

  5. Place into pH 7.01 solution. Leave it there for at least 5 minutes.

  6. Start calibration procedure for pH 7 solution.

  7. Complete the calibration.

After calibration please test accuracy.

Leave pH probe for 5 minutes in pH 4.01 solution and test accuracy. The difference shouldn’t be higher than 0.05.

Then leave the pH probe in pH 7.01 solution for 5 minutes to check the pH probe is accurate to the measurement solution. The difference shouldn’t be higher than 0.05.

Here is the video about the calibration:

Regards

Reef Factory Team