Calibration process for KH keeper's aquarium pump

Calibration process for KH keeper's aquarium pump

To calibrate the aquarium pump please follow these steps:

  1. Make sure the plastic filter is submerged in water.

  2. Put the exit tube coming from the pump to the measuring beaker.

  3. Fully fill the tubes using Fill the circuit button.

  4. Please empty the measuring beaker if any water is present.

  5. Start the calibration process.

  6. When the pump stops please measure the amount that it is in the beaker and put the value in the box.

Check if the pump is calibrated correctly.

To do this put the tube into the empty measuring cup and use Test the accuracy button. The difference should not be greater than 0,5 ml.

Here is the video about the calibration:

Best regards

Reef Factory Team