Leave pH probe to soak in saltwater in your glass beaker for at least 30 minutes. Then please do as follows:
Heat the pH 4.01 and pH 7.01 solution to 25.0 Celsius
Place the probe into the pH 4.01 solution. Leave it there for at least 5 minutes
Start calibration procedure
After the 5 minutes, rinse in RO water, pat dry
Place into pH 7.01 solution. Leave it there for at least 5 minutes
Start calibration procedure
Complete the calibration
After calibration please test accuracy.
Leave pH probe fo 5 minutes in pH 4.01 solution and test accuracy. Difference shouldn’t be higher than 0.05.
Then leave the pH probe in pH 7.01 solution for 5 minutes to check the pH probe is accurate to the measurement solution. Difference shouldn’t be higher than 0.05.