Smart tester - User Manual.


Table of Contents:


I. Set composition.

II. Assembly of the device.

III. Adding the device to the Smart Reef.

IV. Calibrating the measuring cuvette.

V. Installing the cartridges.

VI. Configuration of the measurement schedule.

VII. Resetting the device.

VIII. Updating the device.

IX. Additional functions

X. Automatic actions

XI. Technical information and warnings.

XII. Warranty conditions.

IMPORTANT INFORMATION! Before configuring and using the device, read the entire manual very carefully. There may be slight differences in content between what is presented on the screen of the device and the descriptions in the manual.


I. Set composition.

image-20241110-140428.png

  1. Main unit

  2. Filter holder A

  3. Filter holder B

  4. Calibration scale

  5. Drip pan mounting tool

  6. DIN rail

  7. RO water cartridge

  8. Water intake filter for Ca/Mg tests

  9. Hose 3/5 mm 1.5 m for water intake

  10. Hose 3/5 mm 1.5 m for water waste

  11. 2 x Hose 3/5 mm 0.2 m for device calibration

  12. Power supply

  13. Handle screw

  14. Handle nut

  15. Handle fixing screw

  16. Mixing tablet

  17. Magnet for resetting the device


II. Assembly of the device.

Start assembling the device by unpacking the contents of the package and preparing the necessary components.

image-20241110-140428.png

In addition to the installation process and the first commissioning, prepare:

  1. 2x AAA type alkaline batteries (not included)

  2. Tools and screws needed to mount the device (not included)

  3. A water dish with a minimum capacity of 20 ml.

  4. Paper towels.

After unpacking the components from the package, follow the installation steps shown in the graphics. To mount the device using the dedicated DIN glass, follow the graphic (2).

The kit does not include mounting screws, which must be selected independently depending on the substrate.

1. Connect the power supply cord to the device. At this stage, do not plug the power adapter into an electrical outlet yet.

image-20241110-141150.png

2. Mount the device in the target location. The DIN rail must be mounted horizontally, and the wall on which the device will be mounted must be smooth and vertical. Otherwise, an incorrect position of the device will prevent correct measurements.

image-20241110-143405.png

To hang the device, hook it with the top teeth on the DIN rail, and then press down on the bottom part of the device so that the latches pop into place. Once mounted, you should hear a distinctive click, and the device will hold firmly in the bracket.

CAUTION: Once the device is in operation, never disassemble or move the device without first drying the vessel and hoses, as well as removing the cartridges.

Moving the device, which may contain liquid, risks permanent damage!

To remove the device from the DIN rail, tilt the bottom of the device away from the wall at about 30 - 45 degrees, then lift to release the device from the bracket.

3. Assemble the water filter holder and place the filter with the hose in the holder.

image-20241110-144458.png

Place the plastic pieces with the flat sides facing each other, and then screw them together, using screws. Remember to orient the flat parts of the sump according to the instructions on the graphic. Adjust the height of the bracket so that the top piece hanging on the edge of the sump reaches a depth of about 3 - 5 cm from the bottom. After assembling the holder, place the water filter in it, and slide the hose into the dedicated holder.

4. Place the filter with the handle in the sumptuary.

image-20241110-145226.png

The water intake point for measurement should be behind the mechanical filtration, but before the skimmer.

5. Connect the water connectors to the device.

image-20241110-145436.png

Connect the hose with the water filter to the IN connector, on the left side of the unit. Connect the water waste hose to the OUT connector, with the other end in the water waste container.

Make sure the connector is properly tightened.

Water after measurement, contains substances dangerous to marine organisms and must not be returned to the aquarium under any circumstances.

6. Remove the black cover.

image-20241110-145854.png

Grabbing the sides of the black cover, pull it towards you. The cover is mounted on 4 magnets and has no latches. Removing the cover does not require much force.

7. Place the mixing tablet in the measuring cuvette.

image-20241110-151332.png

Grasp the flat piece on the measuring cuvette and pull it toward you to eject the cuvette. Place the white mixing tablet in the glass dish without touching the measuring cuvette.

If you see any dirt or fingerprints on the glass dish, wipe it off with a paper towel.

Slide the measuring cuvette back into the chamber. Make sure the cuvette is pushed all the way in.

8. Plug the power adapter into a power outlet.

image-20241110-151620.png

Choose the right connector for your region. Then install it in the power supply by twisting it until you hear the characteristic click. Plug the power supply into the power outlet. The LED on the top of the device will light up.


III. Adding a device to Smart Reef.


IV. Calibration of the measuring cuvette.

Przed przystąpieniem do kalibracji kuwety pomiarowej przygotuj potrzebne elementy:

  • 2 x 3/5 mm 0.2 m hose for calibration of the device

  • A vessel with a minimum capacity of 20 ml.

  • 20 - 50 ml of seawater or RO water.

  • Paper towels or a clean cloth.

CAUTION: Do not move the device that contains liquid. Before changing the position of the device, make sure that the measuring cuvette is empty, there are no reagents, RO water container or spilled water in the device.

Place the device on a flat surface. And press Start.

You can calibrate the measuring cuvette at the target mounting location of the device, or in a more convenient location such as a desk. The key factor is a level surface.

Ensure that the device is not tilted to either side when calibrating the measuring cuvette

IMG_0867-portrait-20241110-170044.png

Prepare the necessary elements if you haven't already done so and press Next.

IMG_0868-portrait.png

IMG_0885-portrait.png