  1. Heat the pH 4.01 and pH 7.01 solution to 25.0 Celsius.

  2. Place the probe into the pH 4.01 solution. Leave it there for at least 5 minutes.

  3. Start calibration procedure for pH 4 solution.

  4. After 5 minutes, rinse in RO water, pat dry.

  5. Place into pH 7.01 solution. Leave it there for at least 5 minutes.

  6. Start calibration procedure for pH 7 solution.

  7. Complete the calibration.

...