After calibration please test accuracy.

Leave pH probe fo 5 minutes in pH 4.01 solution and test accuracy. Difference shouldn’t be higher than 0.05.

Then leave the pH probe in pH 7.01 solution for 5 minutes to check the pH probe is accurate to the measurement solution. Difference shouldn’t be higher than 0.05.

Regards