krokstep | opisdescription | zdjęciapicture |
---|
krok step 1 | Remove the measuring cuvette drawer capsule from the Smart Tester. | Image Removed Image Added |
krok step 2 | Take out the magnetic stir bar of the measuring cuvette and put it in a safe place so as not to lose it. | Image Removed Image Added |
krok step 3 | Remove the water supply hoses silicon tubes to the measuring cuvette. Note the order in which the hoses silicon tubes are connected. | Image Removed Image Added |
krok step 4 | Wash the measuring cuvette thoroughly under running water. Cleaning products can be used, but the measuring cuvette should be thoroughly rinsed afterwards. | Image Removed | krok . Attention! If the measuring cuvette is very dirty and it can't be washed with ordinary cleaning products, then use cleaning product based on chlorine or its derivatives, such as sodium hypochlorite. Fill the tray with a chlorine-based cleaner or its derivatives, such as potassium hypochlorite, and leave it for 24hrs. | Image Added |
step 5 | Clean the measuring cuvette dry. Be careful that no dirt, smudges or fingerprints remain on the walls. | Image Removed Image Added |
krok step 6 | Connect the hoses silicon tubes to the measuring cuvette. Pay attention to the correct connection of the hosessilicon tubes. Grasp the measuring cuvette as shown in the photo to avoid leaving fingerprints on it. | Image Removed Image Added |
krok step 7 | Settle the measuring cuvette in the drawercapsule. Don't forget to place the magnetic stir bar in the measuring cuvette. | Image Removed | krok For more details on the correct installation of the measuring cuvette, see this document, please: installation of measuring cuvette | Image Added |
step 8 | Place the measuring cuvette drawer capsule in the Smart Tester. | Image Removed |