step | description | picture |
---|
step 1 | Remove the measuring cuvette capsule from the Smart Tester. |  |
step 2 | Take out the magnetic stir bar of the measuring cuvette and put it in a safe place so as not to lose it. |  |
step 3 | Remove the water supply silicon tubes to the measuring cuvette. Note the order in which the silicon tubes are connected. |  |
step 4 | Wash the measuring cuvette thoroughly under running water. Cleaning products can be used, but the measuring cuvette should be thoroughly rinsed afterwards. Attention! If the measuring cuvette is very dirty and it can't be washed with ordinary cleaning products, then use cleaning product based on chlorine or its derivatives, such as sodium hypochlorite. Fill the tray with a chlorine-based cleaner or its derivatives, such as potassium hypochlorite, and leave it for 24hrs. |  |
step 5 | Clean the measuring cuvette dry. Be careful that no dirt, smudges or fingerprints remain on the walls. |  |
step 6 | Connect the silicon tubes to the measuring cuvette. Pay attention to the correct connection of the silicon tubes. Grasp the measuring cuvette as shown in the photo to avoid leaving fingerprints on it. |  |
step 7 | Settle the measuring cuvette in the capsule. Don't forget to place the magnetic stir bar in the measuring cuvette. For more details on the correct installation of the measuring cuvette, see this document, please: installation of measuring cuvette |  |
step 8 | Place the measuring cuvette capsule in the Smart Tester. For more details on the correct installation of the measuring cuvette, see this document, please: installation of measuring cuvette |  |