  1. Make sure the difference in height between the KH Keeper keeper and the reagent container is not greater than 20 cm.

  2. Put the reagent nozzle coming from the pump to the measuring beaker.

  3. Fully fill the tubes using

    Status
    colourYellow
    titleFill the circuit
    button.

  4. Please empty the measuring beaker if any liquid is present.

  5. Start the calibration process.

  6. When the pump stops please measure the amount of liquid that it is in the beaker and put the value in the box.

Info

Check if the pump is calibrated correctly.

To do this put the tube into the empty measuring cup and use

Status
colourBlue
titleTest the accuracy
button. The difference should not be greater than 0,05 ml.

...