V. Pump calibration.Pump calibration is for the correct calculation of topped-up water statistics. Calibration should be repeated periodically depending on the frequency of operation of the unit. The recommended calibration interval is 1 month. Dosing pump ATO can be calibrated in 2 ways: By volume using a measuring vessel By weight using a jeweller's balance NOTE: Before starting calibration, the device must be placed in the target location, connected to the Smart reef account, and hoses of the target length must be used for calibration.
To perform the calibration you will need:
NOTE: Before the next step, make sure that the water intake hose is in the container from which you want to take water, and that the hose pouring the liquid is over the empty vessel. Then use the "Fill circuit" function to bleed the system.
After the calibration is completed, an accuracy test is recommended to check the correctness of the calibration. Attention: Remember to put the hose back in the target holder. |
...
English | ||||||||
---|---|---|---|---|---|---|---|---|
Accuracy test.The accuracy test is used to verify that the calibration has been carried out correctly. It should be performed after each calibration, and if the values differ significantly, the calibration should be repeated.
Check the amount of fluid pumped. If it is outside your acceptable range, repeat the calibration. Attention: Remember to put the hose back in the target holder. |
...