can’t clean dropper of reagent 1

The error may occurs when:

  • chemical reagent cartidge is empty or not connected rigth to the reagnet pump,

  • chemical reagent dropper or pipes are clogged,

  • chemical reagent dropper not properly seated in its socket, has been bent or damaged,

  • chemical reagent pump is damaged.

Before repeating measurement:

If the problem still occurs, contact Technical Support. Share error code.

...

English us

can’t clean dropper of reagent 2

The error may occurs when:

  • chemical reagent cartidge is empty or not connected rigth to the reagnet pump,

  • chemical reagent dropper or pipes are clogged,

  • chemical reagent dropper not properly seated in its socket, has been bent or damaged,

  • chemical reagent pump is damaged.

Before repeating measurement:

  • be sure reganet still is availabe in cartidge or cartridge is right connected,

  • clean the dropper manualy,

  • make sure the measuring cuvette is installed correctly and there is no dirt in the measuring chamber (read this, please installation of measuring cuvette ),

  • perform reagent change procedure, hovewer use the old reagents as new.

If the problem still occurs, contact Technical Support. Share error code.

...

English us

can’t clean dropper of reagent 3

The error may occurs when:

  • chemical reagent cartidge is empty or not connected rigth to the reagnet pump,

  • chemical reagent dropper or pipes are clogged,

  • chemical reagent dropper not properly seated in its socket, has been bent or damaged,

  • chemical reagent pump is damaged.

Before repeating measurement:

  • be sure reganet still is availabe in cartidge or cartridge is right connected,

  • clean the dropper manualy,

  • make sure the measuring cuvette is installed correctly and there is no dirt in the measuring chamber (read this, please installation of measuring cuvette ),

  • perform reagent change procedure, hovewer use the old reagents as new.

If the problem still occurs, contact Technical Support. Share error code.

...

English us

can’t add reagent 1

The error may occurs when:

  • chemical reagent cartidge is empty or not connected rigth to the reagnet pump,

  • chemical reagent dropper or pipes are clogged,

  • chemical reagent dropper not properly seated in its socket, has been bent or damaged,

  • chemical reagent pump is damaged.

Before repeating measurement:

  • be sure reganet still is availabe in cartidge or cartridge is right connected,

  • clean the dropper manualy,

  • make sure the measuring cuvette is installed correctly and there is no dirt in the measuring chamber (read this, please installation of measuring cuvette ),

  • perform reagent change procedure, hovewer use the old reagents as new.

If the problem still occurs, contact Technical Support. Share error code.

...

English us

can’t add reagent 2

The error may occurs when:

  • chemical reagent cartidge is empty or not connected rigth to the reagnet pump,

  • chemical reagent dropper or pipes are clogged,

  • chemical reagent dropper not properly seated in its socket, has been bent or damaged,

  • chemical reagent pump is damaged.

Before repeating measurement:

  • be sure reganet still is availabe in cartidge or cartridge is right connected,

  • clean the dropper manualy,

  • make sure the measuring cuvette is installed correctly and there is no dirt in the measuring chamber (read this, please installation of measuring cuvette ),

  • perform reagent change procedure, hovewer use the old reagents as new.

If the problem still occurs, contact Technical Support. Share error code.

...

English us

can’t add reagent 3

The error may occurs when:

  • chemical reagent cartidge is empty or not connected rigth to the reagnet pump,

  • chemical reagent dropper or pipes are clogged,

  • chemical reagent dropper not properly seated in its socket, has been bent or damaged,

  • chemical reagent pump is damaged.

Before repeating measurement:

  • be sure reganet still is availabe in cartidge or cartridge is right connected,

  • clean the dropper manualy,

  • make sure the measuring cuvette is installed correctly and there is no dirt in the measuring chamber (read this, please installation of measuring cuvette ),

  • perform reagent change procedure, hovewer use the old reagents as new.

If the problem still occurs, contact Technical Support. Share error code.

...