#error100

English us

“measurement couvete couvette has not been calibrated”

The error may occurs when:

  • not proper calibration factors of measurement couvete has set,

  • measurement couvete has not been calibrated yet.

Before repeating measurement:

  • Ensure that the calibration process of the measuring cuvette has been carried out and completed successfully.

  • Check if calibration factors of the measurement couvette are set. In the Statistics tab, under the items Volume A and Volume B, the calibration coefficients are showed. Typical values of item Volume A is in range 3.5 - 4.5ml and item Volume B in range 0 - 1ml.

  • If nessesery, repeat calibration of the measurement couvete.

If the problem still occurs, contact Technical Support. Share Volume A and Volume B items values and error code.

Polish

"kuweta pomiarowa nie została skalibrowana"

Błąd może wystąpić, gdy

  • nie ustawiono odpowiednich współczynników kalibracji czujnika pomiarowego,

  • czujnik pomiarowy nie został jeszcze skalibrowany.

Przed powtórzeniem pomiaru:

  • Upewnij się, że proces kalibracji kuwety pomiarowej został przeprowadzony i zakończony pomyślnie.

  • Sprawdź, czy współczynniki kalibracji kuwety pomiarowej są ustawione. W zakładce Statistics, w pozycjach Volume A i Volume B, wyświetlane są współczynniki kalibracji. Typowe wartości pozycji Objętość A mieszczą się w zakresie 3,5 - 4,5 ml, a pozycji Objętość B w zakresie 0 - 1 ml.

Image Added
  • W razie potrzeby powtórzyć kalibrację kuwety pomiarowej.

Jeśli problem nadal występuje, skontaktuj się z działem pomocy technicznej. Udostępnij wartości pozycji Volume A i Volume B oraz kod błędu.

#error110

English us

general water issue

If the problem still occurs, contact Technical Support.

Polish

"ogólny problem z wodą"

Jeśli problem nadal występuje, skontaktuj się z pomocą techniczną.

#error111

English us

water level sensor issue

The error may occurs when:

  • photometry circuit is damaged,

  • measuring couvete couvette is dirty.

Before repeating measurement:

  • clean measurement couvette.

If the problem still occurs, contact Technical Support. Share error code.

Polish

"Problem z czujnikiem poziomu wody"

Błąd może wystąpić, gdy

  • obwód fotometrii jest uszkodzony,

  • kuweta pomiarowa jest zabrudzona.

Przed powtórzeniem pomiaru:

  • wyczyść kuwetę pomiarową.

Jeśli problem nadal występuje, skontaktuj się z pomocą techniczną. Udostępnij kod błędu.

#error112

English us

can’t take water sample

The error may occurs when:

  • filling pipe isn’t is not placed in the aquarium water

  • the filling pipe is to long (longer than 2m),

  • the Smart Tester is placed too high in relation to the tank from which the water sample is taken,

  • the hoses are clogged or damaged,

  • filling pump is damaged,

  • measuring couvette is dirty.

Before repeating measurement:

  • be sure the hoses are connected right - hoses to the aquarium water connection - IN and the used water connection - OUT on the Smart Tester are connected as for normal use.,

  • clean measurement couvette,

  • be sure Smart Tester and hoses are connected without any issues and are not coggled,.

If the problem still occurs, contact Technical Support. Share error code.

Polish

"Nie można pobrać próbki wody"

Błąd może wystąpić, gdy

  • rura napełniająca nie jest umieszczona w wodzie akwariowej

  • rura napełniająca jest zbyt długa (dłuższa niż 2 m),

  • Smart Tester jest umieszczony zbyt wysoko w stosunku do zbiornika, z którego pobierana jest próbka wody,

  • węże są zatkane lub uszkodzone,

  • pompa napełniająca jest uszkodzona,

  • kuweta pomiarowa jest zabrudzona.

Przed powtórzeniem pomiaru:

  • upewnić się, że węże są podłączone prawidłowo - węże do przyłącza wody akwariowej - IN i przyłącza wody zużytej - OUT na Smart Testerze są połączone jak przy normalnym użytkowaniu.

  • wyczyść kuwetę pomiarową,

  • upewnić się, że Smart Tester i węże są prawidłowo podłączone i nie są zablokowane.

Jeśli problem nadal występuje, skontaktuj się z pomocą techniczną. Udostępnij kod błędu.

#error113

English us

can’t remove water sample

The error may occurs when:

  • draining pump is damaged

  • measuring couvete couvette is dirty

Before repeating measurement:

  • clean measurement couvette,

  • be sure Smart Tester and hoses are connected without any issues and are not coggled.

If the problem still occurs, contact Technical Support. Share error code.

Polish

"Nie można usunąć próbki wody"

Błąd może wystąpić, gdy

  • pompa spustowa jest uszkodzona,

  • kuweta pomiarowa jest zabrudzona.

Przed powtórzeniem pomiaru:

  • wyczyścić kuwetę pomiarową,

  • upewnić się, że Smart Tester i węże są prawidłowo podłączone i nie są zablokowane.

Jeśli problem nadal występuje, skontaktuj się z pomocą techniczną. Udostępnij kod błędu.

#error120

English us

drops detection circuit don’t work

The error may occurs when:

  • the drop detection circuit is obscured by incorrect placement of the measuring cuvette,

  • drops detection circuit is damaged.

Before repeating the measurement:

  • make sure the measuring cuvette is installed correctly and there is no dirt in the measuring chamber.

If the problem still occurs, contact Technical Support. Share error code.

Polish

"obwód wykrywania kropli nie działa"

Błąd może wystąpić, gdy:

  • obwód wykrywania kropli jest przesłonięty nieprawidłowym umieszczeniem kuwety pomiarowej,

  • obwód wykrywania kropli jest uszkodzony.

Przed powtórzeniem pomiaru:

  • upewnij się, że kuweta pomiarowa jest poprawnie zamontowana i nie występują żadne zabrudzenia w komorze pomiarowej

Jeśli problem nadal występuje, skontaktuj się z pomocą techniczną. Udostępnij kod błędu.

#error121

English us

can’t clean dropper of reagent 1

The error may occurs when:

  • chemical reagent cartidge is empty or not connected rigth to the reagnet pump,

  • dropper or pipes are clogged,

  • chemical reagent pump is damaged.

Before repeating measurement:

  • be sure reganet still is availabe in cartidge or cartridge is right connected,

  • clean the dropper manualy,

  • make sure the measuring cuvette is installed correctly and there is no dirt in the measuring chamber,

  • perform reagent change procedure, hovewer use the old reagents as new.

If the problem still occurs, contact Technical Support. Share error code.

Polish

"Nie można wyczyścić zakraplacza odczynnika 1"

Błąd może wystąpić, gdy:

  • wkład na odczynnik chemiczny jest pusty, albo wkład nie jest prawidłowo podłączony,

  • zakraplacz lub wężyki reagentów są zatkane,

  • pompa odczynnika chemicznego jest uszkodzona.

Przed powtórzeniem pomiaru:

  • upewnić się, że odczynnik jest nadal dostępny w pojemniku, lub pojemnik jest prawidłowo podłączony,

  • ręcznie wyczyścić zakraplacz,

  • upewnij się, że kuweta pomiarowa jest poprawnie zamontowana i nie występują żadne zabrudzenia w komorze pomiarowej,

  • jeżeli to konieczne to przeprowadź procedurę wymiany odczynnika, jednak użyj starych odczynników jako nowych.

Jeśli problem nadal występuje, skontaktuj się z pomocą techniczną. Udostępnij kod błędu.

#error122

English us

can’t clean dropper of reagent 2

The error may occurs when:

  • chemical reagent cartidge is empty or not connected rigth to the reagnet pump,

  • dropper or pipes are clogged,

  • chemical reagent pump is damaged.

Before repeating measurement:

  • be sure reganet still is availabe in cartidge or cartridge is right connected,

  • clean the dropper manualy,

  • make sure the measuring cuvette is installed correctly and there is no dirt in the measuring chamber,

  • perform reagent change procedure, hovewer use the old reagents as new.

If the problem still occurs, contact Technical Support. Share error code.

Polish

"Nie można wyczyścić zakraplacza odczynnika 2"

Błąd może wystąpić, gdy:

  • wkład na odczynnik chemiczny jest pusty, albo wkład nie jest prawidłowo podłączony,

  • zakraplacz lub wężyki reagentów są zatkane,

  • pompa odczynnika chemicznego jest uszkodzona.

Przed powtórzeniem pomiaru:

  • wkład na odczynnik chemiczny jest pusty, albo wkład nie jest prawidłowo podłączony

  • ręcznie wyczyścić zakraplacz,

  • upewnij się, że kuweta pomiarowa jest poprawnie zamontowana i nie występują żadne zabrudzenia w komorze pomiarowej,

  • jeżeli to konieczne to przeprowadź procedurę wymiany odczynnika, jednak użyj starych odczynników jako nowych.

Jeśli problem nadal występuje, skontaktuj się z pomocą techniczną. Udostępnij kod błędu.

#error123

English us

can’t clean dropper of reagent 3

The error may occurs when:

  • chemical reagent cartidge is empty or not connected rigth to the reagnet pump,

  • dropper or pipes are clogged,

  • chemical reagent pump is damaged.

Before repeating measurement:

  • be sure reganet still is availabe in cartidge or cartridge is right connected,

  • clean the dropper manualy,

  • make sure the measuring cuvette is installed correctly and there is no dirt in the measuring chamber,

  • perform reagent change procedure, hovewer use the old reagents as new.

If the problem still occurs, contact Technical Support. Share error code.

Polish

"Nie można wyczyścić zakraplacza odczynnika 3"

Błąd może wystąpić, gdy:

  • wkład na odczynnik chemiczny jest pusty, albo wkład nie jest prawidłowo podłączony,

  • zakraplacz lub wężyki reagentów są zatkane,

  • pompa odczynnika chemicznego jest uszkodzona.

Przed powtórzeniem pomiaru:

  • wkład na odczynnik chemiczny jest pusty, albo wkład nie jest prawidłowo podłączony

  • ręcznie wyczyścić zakraplacz,

  • upewnij się, że kuweta pomiarowa jest poprawnie zamontowana i nie występują żadne zabrudzenia w komorze pomiarowej,

  • jeżeli to konieczne to przeprowadź procedurę wymiany odczynnika, jednak użyj starych odczynników jako nowych.

Jeśli problem nadal występuje, skontaktuj się z pomocą techniczną. Udostępnij kod błędu.

#error124

English us

can’t move back reagents to their cartidges

The error may occurs when:

  • drops of reagent fell when pump stop,

  • reagents hoses are reversed in at least one pump.

Before repeating measurement:

  • clean the dropper manualy,

If the problem still occurs, contact Technical Support. Share error code.

Polish

"Nie można cofną odczynników z powrotem do ich wkładów"

Błąd może wystąpić, gdy:

  • krople odczynnika spadły po zatrzymaniu pompy,

  • węże odczynników są odwrócone w co najmniej jednej pompie.

Przed powtórzeniem pomiaru:

  • ręcznie wyczyść zakraplacz,

Jeśli problem nadal występuje, skontaktuj się z pomocą techniczną. Udostępnij kod błędu.

#error210

English us

general water issue

If the problem occurs, contact Technical Support.

Polish

"ogólny problem z wodą"

Jeśli problem nadal występuje, skontaktuj się z pomocą techniczną.

#error 211

English us

water level sensor issue

The error may occurs when:

  • photometry circuit is damaged,

  • measuring couvete is dirty.

Before repeating measurement:

  • clean measurement couvette.

If the problem occurs, contact Technical Support. Share error code.

Polish

"Problem z czujnikiem poziomu wody"

Błąd może wystąpić, gdy

  • obwód fotometrii jest uszkodzony,

  • kuweta pomiarowa jest zabrudzona.

Przed powtórzeniem pomiaru:

  • wyczyść kuwetę pomiarową.

Jeśli problem nadal występuje, skontaktuj się z pomocą techniczną. Udostępnij kod błędu.

#error212

English us

can’t take water sample

The error may occurs when:

  • filling pipe isn’t is not placed in the aquarium water

  • the filling pipe is to long (longer than 2m),

  • the Smart Tester is placed too high in relation to the tank from which the water sample is taken,

  • the hoses are clogged or damaged,

  • filling pump is damaged,

  • measuring couvette is dirty.

Before repeating measurement:

  • be sure the hoses are connected right - hoses to the aquarium water connection - IN and the used water connection - OUT on the Smart Tester are connected as for normal use.,

  • clean measurement couvette,

  • be sure Smart Tester and hoses are connected without any issues and are not coggled,

If the problem occurs, contact Technical Support. Share error code.

Polish

"Nie można pobrać próbki wody"

Błąd może wystąpić, gdy

  • rura napełniająca nie jest umieszczona w wodzie akwariowej

  • rura napełniająca jest zbyt długa (dłuższa niż 2 m),

  • Smart Tester jest umieszczony zbyt wysoko w stosunku do zbiornika, z którego pobierana jest próbka wody,

  • węże są zatkane lub uszkodzone,

  • pompa napełniająca jest uszkodzona,

  • kuweta pomiarowa jest zabrudzona.

Przed powtórzeniem pomiaru:

  • upewnić się, że węże są podłączone prawidłowo - węże do przyłącza wody akwariowej - IN i przyłącza wody zużytej - OUT na Smart Testerze są połączone jak przy normalnym użytkowaniu.

  • wyczyść kuwetę pomiarową,

  • upewnić się, że Smart Tester i węże są prawidłowo podłączone i nie są zablokowane.

Jeśli problem nadal występuje, skontaktuj się z pomocą techniczną. Udostępnij kod błędu.

#error213

English us

can’t remove water sample

The error may occurs when:

  • draining pump is damaged,

  • measuring couvete couvette is dirty.

Before repeating measurement:

  • clean measurement couvette,

  • be sure Smart Tester and hoses are connected without any issues and are not coggled.

If the problem still occurs, contact Technical Support. Share error code.

Polish

"Nie można usunąć próbki wody"

Błąd może wystąpić, gdy

  • pompa spustowa jest uszkodzona,

  • kuweta pomiarowa jest zabrudzona.

Przed powtórzeniem pomiaru:

  • wyczyścić kuwetę pomiarową,

  • upewnić się, że Smart Tester i węże są prawidłowo podłączone i nie są zablokowane.

Jeśli problem nadal występuje, skontaktuj się z pomocą techniczną. Udostępnij kod błędu.

#error300

English us

photometry circuit don’t work

The error may occurs when:

  • photometry circuit is damaged,

  • measuring couvete couvette is dirty.

Before repeating measurement:

  • clean measurement couvette.

If the problem occurs, contact Technical Support. Share error code.

Polish

"obwód fotometrii nie działa"

Błąd może wystąpić, gdy:

  • obwód fotometrii jest uszkodzony,

  • kuweta pomiarowa jest zabrudzona.

Przed powtórzeniem pomiaru:

  • wyczyść kuwetę pomiarową.

Jeśli problem wystąpi, skontaktuj się z pomocą techniczną. Udostępnij kod błędu.

#error321

English us

can’t add reagent 1

The error may occurs when:

  • chemical reagent cartidge is empty or not connected rigth to the reagnet pump,

  • dropper or pipes are clogged,

  • chemical reagent pump is damaged.

Before repeating measurement:

  • be sure reganet still is availabe in cartidge or cartridge is right connected,

  • clean the dropper manualy,

  • make sure the measuring cuvette is installed correctly and there is no dirt in the measuring chamber,

  • perform reagent change procedure, hovewer use the old reagents as new.

If the problem still occurs, contact Technical Support. Share error code.

Polish

Nie można dodać reagentu 1

The error may occurs when:

  • chemical reagent cartidge is empty or not connected rigth to the reagnet pump,

  • dropper or pipes are clogged,

  • chemical reagent pump is damaged.

Before repeating measurement:

  • be sure reganet still is availabe in cartidge or cartridge is right connected,

  • clean the dropper manualy,

  • make sure the measuring cuvette is installed correctly and there is no dirt in the measuring chamber,

  • perform reagent change procedure, hovewer use the old reagents as new.

If the problem still occurs, contact Technical Support. Share error code.

#error322

English us

can’t add reagent 2

The error may occurs when:

  • chemical reagent cartidge is empty or not connected rigth to the reagnet pump,

  • dropper or pipes are clogged,

  • chemical reagent pump is damaged.

Before repeating measurement:

  • be sure reganet still is availabe in cartidge or cartridge is right connected,

  • clean the dropper manualy,

  • make sure the measuring cuvette is installed correctly and there is no dirt in the measuring chamber,

  • perform reagent change procedure, hovewer use the old reagents as new.

If the problem still occurs, contact Technical Support. Share error code.

Polish

Nie można dodać reagentu 2

The error may occurs when:

  • chemical reagent cartidge is empty or not connected rigth to the reagnet pump,

  • dropper or pipes are clogged,

  • chemical reagent pump is damaged.

Before repeating measurement:

  • be sure reganet still is availabe in cartidge or cartridge is right connected,

  • clean the dropper manualy,

  • make sure the measuring cuvette is installed correctly and there is no dirt in the measuring chamber,

  • perform reagent change procedure, hovewer use the old reagents as new.

If the problem still occurs, contact Technical Support. Share error code.

#error323

English us

can’t add reagent 3

The error may occurs when:

  • chemical reagent cartidge is empty or not connected rigth to the reagnet pump,

  • dropper or pipes are clogged,

  • chemical reagent pump is damaged.

Before repeating measurement:

  • be sure reganet still is availabe in cartidge or cartridge is right connected,

  • clean the dropper manualy,

  • make sure the measuring cuvette is installed correctly and there is no dirt in the measuring chamber,

  • perform reagent change procedure, hovewer use the old reagents as new.

If the problem still occurs, contact Technical Support. Share error code.

Polish

Nie można dodać reagentu 3

The error may occurs when:

  • chemical reagent cartidge is empty or not connected rigth to the reagnet pump,

  • dropper or pipes are clogged,

  • chemical reagent pump is damaged.

Before repeating measurement:

  • be sure reganet still is availabe in cartidge or cartridge is right connected,

  • clean the dropper manualy,

  • make sure the measuring cuvette is installed correctly and there is no dirt in the measuring chamber,

  • perform reagent change procedure, hovewer use the old reagents as new.

If the problem still occurs, contact Technical Support. Share error code.

#error410

English us

general water issue

If the problem occurs, contact Technical Support.

Polish

"ogólny problem z wodą"

Jeśli problem nadal występuje, skontaktuj się z pomocą techniczną.

#error411

English us

water level sensor issue

The error may occurs when:

  • photometry circuit is damaged,

  • measuring couvete couvette is dirty.

Before repeating measurement:

  • clean measurement couvette.

If the problem occurs, contact Technical Support. Share error code.

Polish

"Problem z czujnikiem poziomu wody"

Błąd może wystąpić, gdy

  • obwód fotometrii jest uszkodzony,

  • kuweta pomiarowa jest zabrudzona.

Przed powtórzeniem pomiaru:

  • wyczyść kuwetę pomiarową.

Jeśli problem nadal występuje, skontaktuj się z pomocą techniczną. Udostępnij kod błędu.

#error412

English us

can’t take water sample

The error may occurs when:

  • filling pipe isn’t is not placed in the aquarium water

  • the filling pipe is to long (longer than 2m),

  • the Smart Tester is placed too high in relation to the tank from which the water sample is taken,

  • the hoses are clogged or damaged,

  • filling pump is damaged,

  • measuring couvette is dirty.

Before repeating measurement:

  • be sure the hoses are connected right - hoses to the aquarium water connection - IN and the used water connection - OUT on the Smart Tester are connected as for normal use.,

  • clean measurement couvette,

  • be sure Smart Tester and hoses are connected without any issues and are not coggled,

If the problem occurs, contact Technical Support. Share error code.

Polish

"Nie można pobrać próbki wody"

Błąd może wystąpić, gdy

  • rura napełniająca nie jest umieszczona w wodzie akwariowej

  • rura napełniająca jest zbyt długa (dłuższa niż 2 m),

  • Smart Tester jest umieszczony zbyt wysoko w stosunku do zbiornika, z którego pobierana jest próbka wody,

  • węże są zatkane lub uszkodzone,

  • pompa napełniająca jest uszkodzona,

  • kuweta pomiarowa jest zabrudzona.

Przed powtórzeniem pomiaru:

  • upewnić się, że węże są podłączone prawidłowo - węże do przyłącza wody akwariowej - IN i przyłącza wody zużytej - OUT na Smart Testerze są połączone jak przy normalnym użytkowaniu.

  • wyczyść kuwetę pomiarową,

  • upewnić się, że Smart Tester i węże są prawidłowo podłączone i nie są zablokowane.

Jeśli problem nadal występuje, skontaktuj się z pomocą techniczną. Udostępnij kod błędu.

#error413

English us

can’t remove water sample

The error may occurs when:

  • draining pump is damaged

  • measuring couvete couvette is dirty

Before repeating measurement:

  • clean measurement couvette,

  • be sure Smart Tester and hoses are connected without any issues and are not coggled.

If the problem occurs, contact Technical Support. Share error code.

Polish

"Nie można usunąć próbki wody"

Błąd może wystąpić, gdy

  • pompa spustowa jest uszkodzona,

  • kuweta pomiarowa jest zabrudzona.

Przed powtórzeniem pomiaru:

  • wyczyścić kuwetę pomiarową,

  • upewnić się, że Smart Tester i węże są prawidłowo podłączone i nie są zablokowane.

Jeśli problem nadal występuje, skontaktuj się z pomocą techniczną. Udostępnij kod błędu.

#error423

English us

can’t move back reagents to their cartidges

The error may occurs when:

  • drops of reagent fell when pump stop,

  • reagents hoses are reversed in at least one pump.

Before repeating measurement:

  • clean the dropper manualy,

If the problem still occurs, contact Technical Support. Share error code.

Polish

"Nie można cofną odczynników z powrotem do ich wkładów"

Błąd może wystąpić, gdy:

  • krople odczynnika spadły po zatrzymaniu pompy,

  • węże odczynników są odwrócone w co najmniej jednej pompie.

Przed powtórzeniem pomiaru:

  • ręcznie wyczyść zakraplacz,

Jeśli problem nadal występuje, skontaktuj się z pomocą techniczną. Udostępnij kod błędu.