Make sure the plastic filter is submerged in water
Put the exit tube comming form the pump to measuring beaker
Fully fill the tubes using “Fill the circuit” button
Please empty the measuring beaker if any water is present
Start the calibration process
When the pump stops please measure the amount that it is in the beaker and put the numer value in the box
Info |
---|
Check if the pump is calibrated correctly. To do this put the tube to the empty measuring cup and use “Test the accuracy” button. The difference should not be greater than 0,5 ml. |