The manual includes information about the device functions, which contains the software version number 0.5.3. The software version information is always in the lower right corner of the screen under the product serial number. If the software version is older, proceed with device software update to get access to the latest features.
2. UNBOXING
Widget Connector | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|
|
The following items are included with your kH keeper device (see graphic below):
...
Note |
---|
It is recommended to use the shortest possible hoses. Hoses with a diameter of 2/5 mm should not be longer than 150 centimeters, while the reagent hose should not be longer than 75 centimeters. Remember that any change in hose length affects the amount of liquid dispensed and requires recalibration of the device. Please remember to cut the silicone hoses to the correct length before calibrating the Aquarium & Reagent pumps. |
...
. |
...
Widget Connector | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|
|
3. ASSEMBLY AND DISASSEMBLY OF THE MEASURING BEAKER
If you need to clean the device or disassemble the glass beaker, follow the steps below:
...
Note |
---|
WARNING: It is very important to put the beaker right (to the maximum position) against the back wall of the device. If it is not adjusted properly, there is a risk that the magnetic stirrer will not rotate and the Kh measurement will be incorrect. Each time after mounting the beaker in the device, test the operation of the magnetic stirrer by selecting the speed of its rotation in the device functions (stirring function). |
Widget Connector | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|
|